Thermo Scientific
Orion 2295 Phosphate Analyzer
|
3
Electrical
Power requirements
110 watts. Automatic voltage recognition from 100 – 240 VAC,
110 W (VA), 50 – 60 Hz, Single Phase
Connection
Three wire terminal block (18 – 12 AWG) through a 19 mm (3/4")
bulkhead on side of the electronics enclosure
Safety class
I, protective grounding
Fuse rating
250 VAC, 4 A (fast blow)
Relay contacts
Four potential free relays with 250 VAC, 2 A resistive contact ratings
Outputs
Each relay may be configured separately to a list of programmable
alarm functions
Inputs
Two-isolated analog outputs: 0/4-20 mA, max. load 900 ohms
Environmental
Storage temperature range
5 to 45°C (41 to 113°F)
Operating temperature
range
5 to 45°C (41 to 113°F)
Relative humidity
90% at 40°C (104°F), non-condensing
Altitude
Up to 2000 m
Sample Requirements
Inlet pressure to instrument
1 to 5 psig
Sample flow rate
50 to 1000 mL/min with < 100 microns particles size (user provides
filter)
Sample temperature
5 to 50°C (41 to 122°F)
Sample fitting
6 mm (1/4") OD tubing
Drain fitting
9.6 mm (3/8") OD tubing