Terms Definition
54
|
Orion 2295 Phosphate Analyzer
Thermo Scientific
Section
Section
Analysis Label
Six characters long. Alphabets,
Numerals and +, - are allowed.
Cell Wash Flow
Rate
The rate at which the sample is pumped to
wash the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Reagent Flow
Rate
The rate at which the reagent is pumped to
wash the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Stability
Criteria
Defines the stability level of the data before
accepting it. The stability criteria is a
measure of stability of the absorbance level
rate of 0.001 AU per sec as specified by the
number 1 through 10.
1 through 10.
Post Wash Event
Washing the cell after completion of the
event selected with the following source,
volume and stir time.
Analysis, Calibration, Validation, Not
Assigned.
Post Wash Cycle
Source
Fluid for post washing the reaction cell.
Sample, Reagent, Cleaner, Standard
Post Wash Cycle
Volume
The amount of fluid pumped for post
washing.
-99.9 to +99.9 mL
Post Wash Cycle
Stir Time
The amount of time the fluid is stirred for
post washing the cell.
0 to 999 seconds
CLEANER PARAMETERS
Cleaner Source
Fluid for cleaning the cell.
Sample, Reagent, Cleaner, Standard,
Not Assigned
Volume to add
The amount pumped for cleaning.
0 to 255 mL
Clean Flow Rate
The rate at which the cleaner source is
pumped to clean the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Stir Time
The amount of time the fluid is stirred for
cleaning the cell.
0 to 999 seconds
Frequency
The time duration between two cleaning
cycles.
Manual, Daily, Weekly, Monthly
Start Day/Date
For weekly cleaning the day of cleaning. For
monthly cleaning the date of cleaning.
Day - Week days
Date - Current month dates (1 to 31)
Start Time
The time to start the cleaning on the day or
date selected.
00:00 hr to 23:59 hr
TEMPERATURE UNITS
Units
Units of temperature.
Celsius, Fahrenheit
METHOD PARAMETERS
Cell Wash
Volume
The amount of sample pumped to wash the
cell.
1 to 100 mL
Reaction
Temperature
Control
PID temperature control.
Enable, Disable
Terms
Description
Possible Values, Options