Calibrating Thermistor Messages
Messages
Cause
Solution
Temperature below set
point please check for
hardware malfunction.
May be due to the connectors or
the thermistor or heater
malfunctioning.
Check for thermistor and heater
connections, and calibrate them.
Temperature above set
point please check for
hardware malfunction.
May be due to the connectors or
the thermistor or heater
malfunctioning.
Check for thermistor and heater
connections, and calibrate them.
Measured temperature
differs widely from set
temperature. Please check
for hardware malfunction.
May be due to the connectors or
the thermistor or heater
malfunctioning.
Check for thermistor and heater
connections, and calibrate if
necessary.
Catalog Number
Description
Maintenance and Service Items
2295MK
Maintenance Kit
2230PA
Pump Head Tubing Covers (3 each)
2230PS
Power Supply Assy
2230FS
Power Fuses
2230PI
Air Purge Fittings
2295TK
Main Tubing Kit
2230BC
Empty Auto Cal Btl (Validation Sample Bottle)
2230SF
Fittings, Sample In, Sample Out and Drain
2230RC
Reagent and Standard Bottle Cap - includes fittings
Reagents and Standards
2295REC
Reagent
229510
10 ppm Calibration Standard (For calibration range 0.2-20 ppm)
229530
30 ppm Calibration Standard (For calibration range 20-40 ppm)
229550
50 ppm Calibration Standard (For calibration range 40-50 ppm)