Menu Overview
46
|
Orion 2295 Phosphate Analyzer
Thermo Scientific
Section
Section
Instrument Details Menu
Allows viewing of instrument details:
Software version
Serial number
Model number
Analysis Label Change Menu
Allows technician to edit the analysis label.
Printing Options Menu
Allows technician to print the system configuration, analysis parameters and method
parameters. Also allows technician to enable/disable online printing. See Section 8 for the
default analyzer parameters listing printouts.
Hardware Setup Menu
Calibrate Thermisto
r
Allows technician to calibrate the thermistor. Only required if the factory-installed thermistor has
been serviced or replaced. To calibrate the thermistor, user is required to enter the reaction
block temperature using an external reference thermometer.
Calibrate Pumps
Allows technician to calibrate the flow rate of the pumps.
Calibrate Optics
Allows technician to calibrate the optics.
Calibrate Analog Outputs
Allows technician to calibrate the two analog outputs.
Edit Calibrated Values
Allows technician to edit the calibrated results for fine-tuning.
Prime Fluidics Menu and New Reagents Blank
Auto Prime All
This menu runs an automatic priming sequence for all reagents, standards and sample based
on the priming volume specified by the user.