Section 17- Troubleshooting
Use the NanoDrop Pedestal Reconditioning Compound (PR-1) as a
rapid means of reconditioning the pedestals when the surface
properties have been compromised and liquid columns break during
measurement.
1.
Open the vial containing PR-1 and use the applicator provided in
the kit to remove a pin-head sized amount of the compound.,
2.
Apply a very thin, even layer of PR-1 to the surface of the upper
and lower pedestals.
3.
Wait 30 seconds for the PR-1 to dry.
4.
Fold a clean, dry laboratory wipe into quarters and remove the
PR-1 by aggressively rubbing the surface of the upper and lower
pedestals until all compound residue is removed. Note: The black
appearance of the removed residue is normal.
The reconditioning process is complete once the laboratory wipe
shows is clean of black residue. To check the effectiveness of the
reconditioning, load a 1 ul aliquot of dH
2
0 onto the lower
measurement pedestals and visually verify that the water “beads” up.
Additional information about the PR-1 kit may be found on our
website.
As an alternative to using the PR-1 Kit, the pedestals may be
reconditioned as follows:
1. Fold a clean dry lab wipe over several times to increase its
thickness.
2. Press the lab wipe firmly down on the lower pedestal and
"buff/rub" very aggressively at least 50 times - the lab wipe will
rip during this procedure and will have to be refolded several
times throughout the procedure. The upper pedestal may also be
buffed but care should be taken not to put too much force on the
upper arm.
If the warning persists and the user visually confirms that the liquid
column is forming, contact your local distributor or Technical Support.
Error: Code 8