Section 16- Calibration Check
5.
Add 1ul of deionized water and select “Blank”.
6.
Before opening the ampoule of CF-1 Calibration Fluid, shake
vigorously to ensure solution is thoroughly mixed. Ensure all
solution is collected in the bottom portion of the ampoule.
7.
Carefully break the neck of the ampoule to open the CF-1
Calibration Fluid.
8.
Follow the on-screen prompts in the Customer Guidance text
box. Using individual 1ul aliquots of the CF-1 Calibration
Fluid, measure 10 replicates. After the 10
th
measurement,
the calibration check results will be displayed on-screen in
the Customer Guidance text box.
9.
If the instrument does not pass the calibration check using
1ul samples, immediately repeat the procedure again (step
6) using 2ul samples.
A table of the Calibration Check tolerances and specifications is
available through the Help menu drop down.
To print a copy of the results for your records, click the “Print
Screen” button. A .JPG of the final results is automatically
archived on the hard drive at:
C:\NanoDrop Data\Calib check
If recalibration is required, please contact your local distributor or
Technical Support.
16-2