Section 12- Protein Bradford
standard curve concentration series using the ‘Load Standard Curve
Set-up’.
Delete Standard Points
The Delete Point button allows the user to delete points by first
selecting the data point to be deleted in the table and then choosing
the ‘Delete Point’ button. If erroneous or non-representative readings
are encountered for a specific standard, all replicates of that standard
are cleared by selecting ‘Delete 1 Standard’. Additionally, all
standards can be deleted at once using the ‘Reset all Standards’
button.
Regular Bradford curve
covers 200-8000 ug/ml. Note
the linear range is 100-1000
ug/ml
A mini-Bradford assay covers
an approximate range of 15-
100 ug/ml.
12-6