System Installation, Set Up and Shut Down
Recommended Initial Parameter Settings
2-8
Mercury System Manual
Thermo Fisher Scientific
Table 2–6
lists the 83
i
normal operational parameters.
Note Table 2–6
is a guideline. All values based on 10
µ
g/m
3
at ~9 LPM.
Eductor pressure is set to 5 psig for wet stack and 15 psig during dry stack
application
.
If the orifice pressure is greater than 1.0 psig during a filter
zero or span, it is possible that the probe inlet is clogged (because the outlet
isolation valve is closed, and if the inlet stinger is clogged, the cal gas
pressurizes the fastloop and pressurizes the orifice).
▲
Normal Operational
Parameters (with 83i)