Introduction
Model 83i
1-6
Mercury System Manual
Thermo Fisher Scientific
The Model 82
i
controls the temperatures of the probe, the heated
umbilical line, and converter (located within the probe). These
temperatures are set and recorded via the Model 80
i
.
The Model 82
i
provides both 220 and 110 AC voltage to the Model 83
i
Probe. The 220 VAC powers the probe’s stinger heater and filter/eductor
heater. The 110 VAC powers the Total Hg converter as well as 4 to 6
probe solenoid valves for cal/zero gas, stinger blowback, filter blowback,
and Hg spike.
The Model 82
i
has a dedicated 220 VAC feed to power the heated
umbilical line.
The Model 83
i
Mercury (Hg) Extraction Probe is configured as one
component of the Thermo Scientific integrated Hg Continuous Emission
Monitoring System (CEMS). The Thermo Scientific Mercury Freedom™
System is comprised of a Hg analyzer (Model 80
i
), a Hg Calibrator (Model
81
i
), a Hg Probe Controller (Model 82
i
), and a Mercury Extraction Probe
(Model 83
i
) along with additional peripheral components, such as a zero
air supply, umbilical, and instrument rack. The Model 83
i
with built-in
diluting probe has been designed specifically to monitor gaseous Hg
emissions from coal-fired power plants.
The system is housed in a stainless steel enclosure and is designed to meet
NEMA 4X specifications. To prevent sample condensation, all key
elements (filter, diluting probe and sample eductor) have been mounted
between heated aluminum blocks that can be maintained at temperatures as
high as 250 °C.
The enclosure also houses an accumulator tank for back purging of the
filter tube (blowback). External to the enclosure is a section with the
electrical connections, solenoids for filter and stinger blowback, cal gas, an
Hg spike, a differential pressure transducer, and an additional pressure
transducer for the pressure at the critical orifice.
A special four-inch adapted mounting flange has been supplied for
installation onto the port of the stack or duct.
Clean dry pressurized (~85 psig) air/nitrogen feeds two electronic pressure
regulators in the Model 82
i
, which adjust and maintain output
pressure
to
the Model 83
i
probe eductor and blowback pneumatics (
Figure 1–4
)
.
Pressurized (~65 psig) Hg-free zero air feeds a third electronic regulator in
the Model 82
i
, which adjusts and maintains dilution air pressure to the
Model 83
i
probe dilution Module. Also contained within the Model 82
i
are three electronic pressure transducers associated with the individual
Electrical Power
Model 83
i
Summary of Contents for Mercury Freedom 80i
Page 1: ...Mercury System Manual Part Number 105933 00 28Jan2008 ...
Page 6: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 42: ......
Page 66: ......
Page 79: ...Optional Equipment Nitrogen Generator Thermo Fisher Scientific Mercury System Manual 6 9 ...
Page 100: ......