System Installation, Set Up and Shut Down
System Checkout
2-4
Mercury System Manual
Thermo Fisher Scientific
Check that the 81
i
gas mode display mirrors what is shown on the 80
i
gas
mode display as follows: 80
i
Gas Mode > Instrument Controls > Gas
Modes >
Analyzer Zero
. Refer to the individual Mercury instruction
manual if displays do not match and units are not communicating.
1.
Inside the 82
i
, ensure that both circuit breakers are
ON
.
2.
At the 80
i
Main Menu, select > Instrument Controls > Component
Power >
Umbilical 1 Power
, turn umbilical 1 heater
ON
,
and
Umbilical 2 Power, if applicable.
3.
At the 80
i
Main Menu, select > Diagnostics > Temperatures > Probe >
Umbilical
, and observe temperature rise to 120 °C. Verify the
temperature control is within 10 °C of set point.
4.
At the 80
i
Main Menu, select > Instrument Controls > Component
Power >
Converter Power
, and turn the converter
ON
.
5.
At the 80
i
Main Menu, select > Diagnostics > Temperatures > Probe >
Converter
, observe the converter temperature rise and control to
760 °C. Verify the temperature control is within 10 °C of set point.
6.
At the 80
i
Main Menu, select > Instrument Controls > Component
Power >
Probe Power
, and turn probe power
ON
.
7.
At the 80
i
, select > Diagnostics > Temperatures >
Probe
and observe
temperature rise to 220 °C. Verify temperature control is within 10 °C
of set point.
8.
At the 80
i
Main Menu, select > Service >
Set Pressures
, and set the
following parameters to the value indicated:
Blowback Pressure
60
psig
Eductor Pressure
15 psig for dry stack 83
i
applications, 5 psig
for wet stack applications, 15 psig for 83
i
GC
Dilution Pressure
55 psig for 83
i
, 35 psig for 83
i
GC
9.
While in Sample mode, at the 80
i
Main Menu, select > Instrument
Controls >
Auto/Manual Mode
, and switch to
Manual Hg(t)
.
System Checkout
Summary of Contents for Mercury Freedom 80i
Page 1: ...Mercury System Manual Part Number 105933 00 28Jan2008 ...
Page 6: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 42: ......
Page 66: ......
Page 79: ...Optional Equipment Nitrogen Generator Thermo Fisher Scientific Mercury System Manual 6 9 ...
Page 100: ......