Troubleshooting
System Troubleshooting Guides
Thermo Fisher Scientific
Mercury System Manual
4-15
Figure 4–1.
Example of Model 80
i
Interface Board Voltage Test Point Locations
LEDs on the 82
i
measurement interface board are used to troubleshoot the
82
i
and the 83
i
. Each LED represents a valve in either the 82
i
or 83
i
.
When an LED is ON, the associated valve is ON. Refer to
Table 4–6
for a
recommended troubleshooting action to perform. In most cases, the
problem will involve wiring and/or plumbing.
Table 4–6.
Probe Troubleshooting – 82
i
Measurement Interface Board LEDs
LED No. Description
Action
4
Blowback stinger (83
i
)
5
Blowback filter (83
i
)
Check air supply and pressures via diagnostics.
Check air at the probe.
6
Cal zero (83
i
)
7
Hg spike (83
i
)
Verify that the 81
i
is outputting cal gas, diagnostics,
flows. Check at probe also.
8 Eductor
(82
i
)
Check failsafe temperature.
9 Dilution
(82
i
)
Check air supply and pressures via diagnostics.
Check air at the probe.
10
Mercury valve (83
i
)
Verify that the 81
i
is outputting cal gas, diagnostics,
Probe Troubleshooting
with 82
i
LEDs
Model 80
i
Only