Introduction
Model 81i
Thermo Fisher Scientific
Mercury System Manual
1-3
(Model 82
i
), and a Hg probe (Model 83
i
), along with additional peripheral
components, such as a zero air supply, umbilical, and instrument rack.
However, the Model 81
i
is also available as a stand-alone calibrator.
Table 1–1
shows the Model 81
i
solenoid valve status for the different
calibrator modes. Use this table along with
Figure 1–2
when reading the
principle of operation description that follows. Note that a valve status
shown as ON describes a valve that is energized, or in the Normally Closed
(NC) state. A valve status shown as OFF describes a valve that is de-
energized, or in the Normally Open (NO) state.
Table 1–1.
Model 81
i
Valve Mode Status
Solenoid ID
Standby
Instrument
Zero
Instrument
Cal
Orifice
Zero
Orifice
Cal
System
Zero
System
Cal
Valve
1 ON ON
ON ON ON ON ON
Valve
2 OFF OFF ON OFF ON OFF ON
Valve
3 OFF OFF ON OFF ON OFF ON
Valve 4
OFF
OFF
OFF
ON
OFF
ON
OFF
Valve 5
OFF
OFF
ON
ON
OFF
ON
OFF
Orifice
Valve
OFF OFF
OFF ON ON OFF OFF
Spike
Valve
OFF OFF
OFF OFF OFF ON ON
Clean, dry, Hg-free pressurized air/nitrogen (30-40 psig) feeds both a high
flow (0-20 slpm) dilution air mass flow controller (MFC) and a low flow
(0-50 sccm) Hg source MFC (
Figure 1–2
).