System Installation, Set Up and Shut Down
Start-Up
Thermo Fisher Scientific
Mercury System Manual
2-3
Example: From the Main Menu, select > Instrument Controls >
Gas Mode >
Instrument Zero
.
b.
Verify that the 81
i
changes to that mode.
4.
To test communication from the 80
i
to the 82
i
:
View Diagnostics from the 83
i
, such as Voltages > Interface Board
82
i
> Temperatures > Probe, or Pressure > Probe.
Verify that readings are not zero. Also, unplug one of the 83
i
thermocouples and verify that the change is indicated on the 80
i
.
Use the following procedure to power up the Mercury CEMS.
1.
Place the System in System Zero Mode to flush the system.
2.
Turn OFF flow related components:
Power OFF the sample pump
Component Power > Eductor Power
Service > Probe Set Temp Failsafe > Set To (250
°
C)
3.
Heat components to temperatures:
Component Power > Umbilical 1 Power
Component Power > Umbilical 2 Power
Component Power > Converter Power
Component Power > Stinger Power
Component Power > Probe Power
Component Power > Oxidizer
4.
Restart flow related items:
Service > Probe Set Temp Failsafe > Set To (normal Set-Point)
Component Power > Eductor Power
Power ON the sample pump
5.
Place the system in Sample Mode.
Start-Up