Calibration
Manual System Calibration
3-2
Mercury System Manual
Thermo Fisher Scientific
12.
From the Main Menu, select Instrument Controls > Auto/Manual
Mode >
Hg(t)
.
13.
Repeat Steps 4 through 11 for Hg(t)
mode.
14.
From the Main Menu, select Instrument Controls > Auto/Manual
Mode >
Hg(0)/Hg(t)
.
15.
Set Average Time to 120 seconds.
16.
From the Main Menu, select Instrument Controls > Gas Mode >
System Zero
.
17.
Wait approximately 15 minutes for the instrument to stabilize.
18.
From the Main Menu, select Calibration >
Cal Hg(0) Background
,
and press
.
19.
From the Main Menu, select Calibration >
Cal Hg(t) Background
,
and press
.
20.
From the Main Menu, select Calibration > Auto Zero/Span Check >
Sys Hg Span Conc
, and choose the target concentration (Span 1
through 6).
Note
The Sys Hg Span Conc screen is available only while in Hg(0) or
Hg(t)
Manual mode.
▲
21.
From the Main Menu, select Instrument Controls > Gas Mode >
System Span
.
22.
Wait approximately 15 minutes for the instrument to stabilize.
23.
From the Main Menu, select Calibration >
Calibrate
Hg(0)
Coefficient,
and enter the 81
i
output concentration pressure.
24.
Repeat for Hg(t) Coefficient.