Deuterium lamp
UV radiation from a deuterium lamp can be harmful to the skin and eyes.
Always view the lamp through protective glasses that will absorb UV
radiation.
1.
Turn off the spectrophotometer and unplug from the AC outlet.
Caution
Allow at least 15 minutes for the instrument to cool before removing
protective cover.
2.
Remove back cover.
3.
Remove the burned out bulb from its socket.
a.
Using a 2.5 mm hex driver, turn screws 2 turns
counterclockwise.
b.
Rotate the lamp and pull upward until out of housing.
c.
Unplug inline connector.
4.
Insert new lamp.
Caution
Never touch the bulb with your fingers. Oil from your skin may cause the
bulb to burn out quickly or explode.
a.
Plug in inline connector.
b.
Place lamp in lamp housing aligning hex screws.
c.
Using 2.5 mm hex driver tighten screws.
5.
Replace rear cover.
6.
Plug in the spectrophotometer to the AC outlet and turn on the
power switch.
Note
Lamp hours and energy (if applicable) must be reset from the controlling
software.
80 Evolution 300 and Evolution 600 User Guide