2.
Highlight each parameter in turn and press
Enter
.
Parameter
Function
Data Mode
Toggles between %T and ABS.
Method Name
Allows user to enter a method name.
Wavelength
Selection*
Allows the user to select either Single, Multiple, or
Serial Wavelength measurements.
Wavelength(s)
Sets the wavelength(s) where the measurement(s)
is/are made.
Lamp Change
Allows the user to select between deuterium,
automated (instrument uses either the deuterium or
tungsten lamp to run the selected wavelength) and
tungsten lamps.
Bandwidth
Sets spectral bandwidth.
Integration
Allows the user to determine the integration
(measurement) time.
Timers
Enables up to four countdown timers to be
incorporated into the method sequence. A timer title
allows the user to specify the action (e.g., shake, boil)
associated with the timer. At the end of the timer
period, the program may continue to the next step or
pause to await action. Timers may be multiple use or
single use.
UV
calc
Refer to “
*
Use the table below to help you choose between the three different wavelength options.
Select this
wavelength…
To do this…
Single Wavelength
Measure each sample at one wavelength (the same
for each sample).
Multi Wavelength
Measure each sample at up to 20 wavelengths (the
same for each sample).
Serial Wavelength
Measure up to 9 samples at one wavelength (different
for each sample).
46 Evolution 300 and Evolution 600 User Guide