Thermo Scientific CryoStar NX70 Operator Guide
87
Optional Foot Pedal Emergency Stop
A second emergency stop device is integrated into the optional foot pedal.
To activate the emergency stop using this method:
•
Firmly step on the foot pedal to immediately stop the motorized cutting drive.
Notes
The button for the handwheel brake does not light up when the foot pedal emergency stop is activated,
however the display will show a red button as per the standard emergency stop method.
This emergency stop device will remain active as long as the foot pedal is engaged.
•
The motorized cutting drive can then be started again.
Cooling Processes
Specimen and blade holder cooling
Integrated refrigeration systems allow the blade holder to be cooled to -35°C and the specimen head to be
cooled to -50°C. Specimen stations are located on the left side inside the Cryochamber.
These specimen stations passively cool specimens to a temperature of -30°C and can be used as a staging
area.
The temperature range of the specimen head is adjustable b10°C and -50°C.
The instrument is also equipped with an active Cryobar.
The Cryobar area actively cools down to -60°C.
Note
Use the heat extractor to accelerate the freezing of the specimen.