Thermo Scientific CryoStar NX70 Operator Guide
20
Location & Setup
When deciding where to locate the CryoStar NX70, the following items must be taken into
consideration:
•
Approximately 10 cm (4 in) must be allowed around the cooling system to allow free air flow.
•
Approximately 10 cm (4 in) must be allowed between the wall and the rear panel.
•
The vents on the sides of the rear panel must be unobstructed at all times.
•
The mains power socket must be accessible at all times.
•
The location must be free from draughts, open doors or air conditioning systems.
•
The location must not expose the cryochamber to direct sunlight.
•
Enough room for height adjustment.
Note
These measures are intended to reduce the build-up of frost and maintain consistent cooling temperatures
resulting in more favourable working conditions. Locations which have high humidity and / or high
ambient temperatures can reduce the performance of the instrument.
Once the CryoStar NX70 has been located, ensure the fixing bolts shown are tightened to
prevent the unit from moving during operation.