Technical Specifications
Mechanical Specifications
Depth
100 cm (39.4 in)
Width
75.5 cm (29.7 in)
Height Range
Adjustable range 82 - 112 cm (32.3 - 44.0 in)
Maximum Weight
200 kg (440 lb)
Electrical Specifications
Power Requirements
Dedicated line recommended.
100V, 50/60Hz, 10A, +/-10%, 110-120V, 60Hz, 10A, +/-10%
220-230V, 50/60Hz, 5A, +/-10%
Environmental Specifications
For Indoor Use Only
Temperature (Operating
Limits)
+5°C to +35°C (+41°F to +95°F)
Temperature
(Recommended
Operation)
+15°C to +30°C (+59°F to +86°F)
Note: Performance may deteriorate when operated outside of this
range.
Temperature
(Transport and Storage)
-20°C to +40°C (-4°F to 122°F)
+70°C (158°F) for short exposure
Relative Humidity
Max. 80% RH up to 31°C
Decreasing linearly to 50% RH at 40°C
Altitude
Up to 2000 m (6,500 ft)
Pollution Degree
2
Over Voltage Category
II
Floor Loading
Requirements
300 kg/m² (660 lb/ft²)
Protection Class
I
Sound Emission
<45 dB (A) - measured with 1m distance to the instrument.
Refrigerant
R404a, filling amount 400g