Multi-Function Button
The Multi-Function button provides to the user additional information on the sectioning process.
•
Four different indications can be displayed by pressing this button.
Remaining Distance to the Front End Position
This button indicates the remaining distance to front end position of the blade holder.
Section Thickness Sum
This button indicates the total section thickness sum of the actual mounted specimen.
•
To reset this value, press this button and keep it pressed for approximately 3 seconds.
Note
It is recommended to reset the section thickness sum if a new specimen is mounted in the specimen head for
sectioning.
Motorized Cutting Drive
The motorized cutting drive facilitates routine work and ensures a steady cutting speed even for hard
specimens.
When using the motorized cutting drive the hand wheel is controlled automatically.