H-Genie
®
User Manual
v. 1.7.
page:24 of 29
Figure 25. Shutdown button
The H-Genie
®
will now start to shutdown. If there is any internal hydrogen present, then
it will be released out the back and through the water vent. This will make a noise, but
do not be alarmed, this is completely normal. The entire procedure takes ~60 seconds.
Once it is finished, the status will
change to “You can turn the device off”.
Figure 26: Shutdown phases.
The same procedure will take place if the emergency button is pressed or the software
triggers an emergency shutdown, such as if there is a cell failure or hydrogen leak
detected. The below screens will appear.
Please note! If you intend to switch back on the instrument straight
away after shutting down, then please wait a minimum of 30 seconds
before switching on.