4.2
Location
•
The H-Genie
®
must
be located on a flat, level surface inside a fully functional
ventilated cabinet or ventilated area, such as a fume hood. This minimizes any
risk related to a hydrogen leak into the environment.
•
The fume hood or ventilated cabinet must be equipped with a standard electrical
socket and a water hole.
•
The oxygen and hydrogen vent tube outlets should be positioned as far apart
as possible within the fume hood.
•
It is recommended that any objects be placed a minimum of 15 cm away from
the rear and the sides of the H-Genie
®
to avoid any obstruction to the airflow in
the cabinet and to allow any warm air from within the generator to be released
freely into the environment. Ensure that none of the rear vents are obstructed.
Failure to ensure this may result in damage to the H-Genie
®
.
•
Dimensions of the H-Genie
®
can be found in the Technical Data section.
•
Please adhere to the operating temperatures below:
o
Minimum Operating Ambient Temperature: 10 °C (50 °F)
o
Maximum Operating Ambient Temperature: 35 °C (95 °F)
Figure 8: Operating requirements