4.3.4. Connect one end of the water gas outlet tube to the water drain outlet and the
other end to a sink or a wastewater receptacle located in the fume hood. Turn
the blue switch to open and watch the water flowing out of the H-Genie
®
. Pay
particular attention to the bubbles in the water stream. Wait until no more air
bubbles appear in the water stream (this may take a minute) and then close the
blue valve switch. Take the tube out of the water drain outlet and place it in the
water vent outlet. Please ensure the open end of this tube is fastened to avoid
movement when hydrogen gas is released. Top up the water tank.
4.3.5. Attach the stainless steel L-shaped hydrogen vent tube to its respective
connection.
Figure 12: Attached hydrogen vent, oxygen vent and water vent tubes.
Use water drain port with tubing from water vent if needed
Do not impede or block the oxygen gas outlet at the top of the water
reservoir while the machine is switched on. Only use the cap provided
with the machine.
Do not attempt to drain any of the water tank unless the H-Genie
®
is