5.1.6 Once the reaction is completed or you have no more need of the H-Genie
®
, then
STOP the H-Genie
®
. Never disconnect the H-Genie
®
from the reactor while the
reactor or tubing is under pressure or the H-Genie
®
is still running.
5.1.7 The H-Genie
®
will stop automatically after 1 hour if no hydrogen consumption is
detected.
5.1.8 Once the H-Genie
®
has stopped providing hydrogen, the system will enter a
“Ready” state. After a period of 1 hour without use, the H-Genie
®
will enter
“Standby” state.
5.1.9 Do not restart the H-Genie
®
during an ongoing reaction. The self-test will
recognize that the external pressure is higher than the internal pressure and the
H-Genie
®
will enter Emergency mode.
5.2
Setting a FLOW reaction
5.2.1 Connect the check-valve to the front of the H-Genie
®
.
Ensure that the arrow
on the check-
valve is pointing “out” from the H-Genie
®
.
This is the direction
of the hydrogen flow.
5.2.2 Connect your flow reactor to the check-valve using the appropriate tube and
connection. Ensure there are no leaks and connections are tight.
5.2.3 Press the FLOW function on the main screen.
Figure 18: Main Screen
5.2.4 Press the
button to enter a reaction pressure and hydrogen gas flow rate
value.