4.3
Installing the H-Genie
®
Open the packaging and remove all the contents.
4.3.1. Place the H-Genie
®
reactor box inside the selected fume hood cabinet.
4.3.2. Remove the temporary black water reservoir cap and fill
the water reservoir with
high-purity de-ionized water. Cover the water reservoir hole with the metal cap
provided. Make sure a
ScavCart™ is screwed into the metal cap. For further
instructions please check “5.5 Using and Changing a ScavCart”.
Figure 9: Water reservoir caps
The H-Genie
®
generates hydrogen through in-situ water electrolysis. The
operation of high-pressure cells requires high purity, de-ionized water. Millipore
Milli-Q
®: <1µS/cm is strongly recommended!!!
Using low quality water can cause irreparable damage of the instrument.
The volume of the water reservoir is an approx. 3500 mL. Take care not to overfill. The
maximum water level should be 2 cm below the water tank neck.
Attention