H-Genie
®
User Manual
v. 1.7.
page:15 of 29
4.3.6. Attach the emergency stop button to the connector at the rear of the H-Genie
®
and ensure that the button is located in an easily accessible place. Ensure that
a green band is showing underneath the red button. If not, then turn the red
button counter clockwise to release the button.
4.3.7. Make sure to switch the main power switch to the
„0” (off) position before
inserting the power cable into the rear of the H-Genie
®
. Connect the power cable
to the mains power supply as well.
5.
Performing a reaction with the H-Genie
®
Safety instructions:
Once the installation is complete, switch the main power switch, located at the rear of
the H-Genie
®
to the „I” (on) position. When the initialisation process is ready to start
t
he following screen will appear. Press „OK” to begin the H-Genie
®
’s selftest.
Important! The hydrogen vent tube and the water vent tube will release
hydrogen into the fume hood. Please ensure that the ends of these
pipes are at least 1.5 m away from the outlet of the oxygen outlet to
ensure hydrogen and oxygen do not mix or, if possible, in separate
fume hoods.
The electric cables must be guarded against moisture.
During use the fume hood MUST be switched on, and the fume hood
sash pulled down.
The H-Genie
®
generates hydrogen internally. In normal operation,
hydrogen will be released from the rear ports. It is essential that
adequate precautions are taken to allow this to be safely ventilated by
a fume hood cabinet.
Ensure that the reactor vessel is connected to the H-Genie
®
and gas
tight. Check to ensure that there are no leaks.
Please ensure the water reservoir is topped up at all times. It is vital
that the electrolysis cell membrane is kept constantly wet. The cell
must not be allowed to run dry or the cell will immediately and
irreversibly cease to function.