User Manual for the FLUXMEDICARE® V2 System
-
IFU-FLUXMEDV2 - V.1.0 - 07/2020
Page 54 / 84
9.
Control of the source
The source must be checked before each use for obvious visible damage, damaged cables, cracks, worn label.
The source is guaranteed for 2 years against material or manufacturing defects, from the delivery date. During
this period, the defective equipment will be replaced free of charge, unless the fault is due to abuse or incorrect
use.
A control of the source must be performed annually by MDB Texinov® personnel.
The light source is guaranteed for 5000 hours’ use.
MDB TEXINOV® will only take responsibility for the safety, reliability and perfect operation of the appliance if:
•
All the adjustments as well as modifications or repairs are carried out by MDB TEXINOV®.
•
The device is used in compliance with the instructions for use supplied.
Important indications
Only MDB TEXINOV® personnel are authorised to repair FLUXMEDICARE® V2 products.
The units must not be opened.
If the equipment must be sent back, it must be packed in its original packaging (see diagram below).
10.
Recycling
Electrical appliances must not be discarded with household waste.