6 Using the product
39
>
[Options]
→
Configure Graphic
: the measurement
parameters to be represented (max. 4) can be displayed (
)
or hidden (
).
>
[Options]
→
Mean
: the mean values of a measurement are
displayed.
>
[Options]
→
Number of lines
: change the number of readings
shown per display page.
>
[Options]
→
repeat
: in the case of the compliance
measurement and approval test, the measurement can be
restarted several times within the first 2 minutes if necessary.
6.3.
Conducting a measurement, testo 330-2 LL/F
(0632 3305)
6.3.1.
Preparing for measurement
The time required for an official measurement is approx. 30 min.
If you switch to wood chips as fuel immediately following
a measurement using wood pellets or split logs as fuel, the
message
Particulate matter sensor and sensor cover
need to cool down (this process may take a few
minutes)
appears. Prior to selecting wood chips as fuel,
Testo recommends opening the particulate matter sensor
cover so that the sensor can cool down, . To shorten the
cooling down period, the sensor should be moved to a cold
environment or placed on a cold surface.
The
First steps
chapter must have been read.
CAUTION
Preventing measurement errors
> Do not place or operate the measurement system on heat
sources, e.g. boilers.
Tightness test
For an official measurement, the tightness of the measurement
system must be checked prior to every measurement.
The locking lever for fixing the testo 330 in place must be
closed.
Call up function:
>
[ ]
→
Measurement options
→
[OK]
→
Particle
→
[OK]
→
Tightness test
→
[OK]
.