6 Using the product
35
>
[Options]
→
Recalibrate
: (this function is not available during
a measurement): The gas sensors are set to zero.
>
[Options]
→
Address/Location
: (this function is not available
during a measurement): the Address/Location folder is opened.
>
[Options]
→
Particle sensor pump: on/off
: the sensor pump
is switched on or off.
>
[Options]
→
Test Bed (gradient check)
(only for test beds
and Testo Service):
procedure for checking the determinant
measurement parameters for the six-monthly inspection of the
test bed with specialised equipment.
>
[Options]
→
Diagnosis (service)
:
only for Testo Service.
Measuring
1. Select compliance measurement, adjustment help or approval
test:
[
▼
]
→
[OK]
.
2. Select solid fuel:
[
▲
]
,
[
▼
]
→
[OK]
.
3. Select parameter:
[
▲
]
,
[
▼
]
→
[Edit]
.
Fuel humidity (±15 %u), ambient temperature (±3 °C
amb.
temp.) and ambient humidity (± 15 %RH
amb. humidity)
influence the measurement result. The values specified are
factory settings and need to be adapted to the existing
conditions. At a measuring location, the ambient
temperature and humidity should always be determined
from the same location.
Testo recommends the testo 606-2 wood moisture
measuring instrument for determining the above parameters
(item no. 0560 6062-2).
When the parameter
Part load
is selected, a signal is
sounded after 5 min when a measurement is conducted.
The signal is very helpful for measurements in the partial
load range as per § 25 para 2 of the 1st BImSchV.
If the measurement system is exposed to ambient
temperatures that are outside the operating temperature
range specified in the technical data, a longer stability time
can be expected.
The temperature of the measurement system should not
differ significantly from the ambient temperature at the
measuring location. If this is the case, allow the
measurement system to adjust to the ambient temperature
prior to operation.