7 Maintaining the product
63
3. Depress the lock and pressure element and remove.
4.
Remove the rotation disc.
5. Insert the brushless side of the cleaning brush into the gas path
6. Push the cleaning brush through to the probe shaft tip.
7.
Pull the cleaning brush out of the probe shaft. Repeat steps 5
and 6 several times depending on the degree of contamination.
8.
Replace the rotation disc (pay attention to the locking pin on the
probe and the protrusion on the rotation disc).
9. Reinsert the pressure element and lock.
10. Close the rotation diluter cover (turn clockwise).
11. Insert the particulate matter probe into the holder and secure
with the rubber ring.
12. Lock the transport securing device.