2 Safety and the environment
8
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your
area when performing the measurements.
> The measurement system has been verified as a short-term
measuring instrument and should not be used as a safety
(alarm) instrument.
> Supervise
the
measurement
system during operation.
> The measuring instrument can be operated in closed rooms.
The maximum allowable concentration (MAC) value for CO
of 30 ppm is attained by operating the measuring instrument
under the following boundary conditions after approx. 4.5 h
measuring time at the earliest (in cases where there is no air
exchange):
Room dimensions
(L x W x H)
3 m x 3 m x 2 m
Flue gas concentration 5000 ppm CO
> Do not store or transport any fluids in the transport case.
> Do not operate the instrument if there is condensation on the
housing.
WARNING
There is water in the transport case.
Risk of electric shock!
> Do not operate the measurement system.
> Check for fluids before each start-up.
> Do not store or transport any conductive incidentals (without
protective cover/container) in the transport case.
> Do not use mains cable if there is any external damage to it and
protect it from hot surfaces.
> Secure measurement system during transport in order to
prevent damage to the measurement system.