AMETEK CTS
NSG 3040A, 3060A series
Operating Manual
V 1.04
73 / 157
8.5.
Power Magnetic Field as per IEC 61000-4-8
Parameter
Value
H-Field
1 to approx. 40 A/m using MFT 30
depend magnetic factor setting
Frequency
50/ 60 Hz depend on mains frequency
Coil factor settable
Settings
Transformer factor [V/A]
0.01 to 100 setting range
Coil factor setting [m
-1
]
0.01 to 100 setting range
typical 0.87 for 1 m x 1 m coil
Protection
PF1: open, not connected by the user
PF2 set to zero volt when not in operation
Magnetic field tests per IEC 61000-4-8 and -9
The test routines for handling the magnetic field tests are included in the internal NSG firmware. All functions to
control external options as voltage/current sources or magnetic field antennas are included. In addition, the following
hardware is required:
Option required for Magnetic Field Test 50/60Hz per IEC 61000-4-8
- External transformer (VAR 3005A-S16) and magnetic field antenna (square 1 m x 1 m coil MFC 1000 or MFC
1000.1)
- External current transformer (MFT 30) with 1 turn to test 1, 3, 10 and 30 A/m levels
- External current transformer (MFT 100) with 1 turn to test 100, 300 and 1.000 A/m levels (300 and 1.000 A/m are
short-term tests of 3 sec.)