NSG 3040A, 3060A series
Operating Manual
V 1.04
115 / 157
14.2.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Explanations
A
Phase angle
td
Duration of a single event
tr
Repetition rate (time between two events)
Mode
Channel select (PF1, PF2 or
V)
Vdip
Variable test voltage (controlled by 0-10 V analog voltage)
n
Number of events
sync
Trigger mode synchronization to mains (On / Off)
The parameter to change can be selected by pressing a function key. The corresponding range will then be
displayed.
14.2.1.1. Voltage DIPS
For testing voltage DIPS select:
Mode
U
: During the test the internal switch toggle the
supply voltage between PF1 and PF2
Vdip
:
Dip voltage [%Vnom] or [V]
td
:
Duration of the dip
tr
:
Repetition time between two dips [ms, s]
Sync
:
Synchronization with
phase angle A
[º]
n
:
Number of events
14.2.1.2. Voltage Interruption
For testing voltage DIPS select:
Mode
PF1
: During the test the internal switch PF1 will
switch ON and OFF during the test
td
:
Duration of the interruption
tr
:
Repetition time between two interruption
Sync
:
Synchronization with
phase angle A
[º]
n
:
Number of events