![Terumo BCT Trima Accel Service Manual Download Page 88](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436088.webp)
•
24 V current sense line (DC voltage) from the motor driver CCA must not trigger the solenoid fault
line
•
Safety CCA must turn off the 64 V switch
•
Lock solenoid power must be enabled
–
Safety system controls the lock solenoid power and turns off the solenoid power if the
centrifuge spins over 60 rpm
–
For safety reasons, it is not possible to have both the solenoid energized and the centrifuge
motor powered up (64 V switched)
–
During boot-up, a T1 door lock test is performed that locks and unlocks the door solenoid
Figure 2-23: Door lock system diagram
Control Computer
Door lock movement is initiated when the device software or STS tells the control computer to
lock or unlock the door. The control computer tells the control CCA field-programmable gate
array (FPGA) to actuate the door lock solenoid and also informs the safety computer what it told
the control CCA FPGA through an Ethernet connection. This is a redundant system used to
ensure that the device fails in a safe state. An alarm is generated if the safety computer notices a
difference between the control and safety conditions.
Door System
Trima Accel
®
Automated Blood Collection System • Service Manual
2-27
Summary of Contents for Trima Accel
Page 1: ...Trima Accel Automated Blood Collection System Service Manual...
Page 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Page 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Page 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Page 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Page 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Page 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Page 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Page 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Page 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Page 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Page 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Page 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Page 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...