Blood Run Collection Substates
The Blood Run state contains nineteen prime and collection substates and eleven recovery substates.
This section describes the prime and collection substates.
Channel Setup Substate
This substate continues to establish the interface until a predetermined inlet volume has been pumped.
The pumps begin ramping up in speed during this substate. After completion of this substate, the Trima
Accel system advances to Pre-Platelet Plasma, Pre-Platelet No Plasma, or Plasma Only substates
depending upon the chosen procedure.
Table 3-30: Channel Setup Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
0*
Qret
Return
Return
Return
3000
* DRBC tubing sets
Exit Condition
When a total of 200 mL is processed by the inlet pump.
Pre-Platelet Plasma (Pre-Platelet, Collecting Plasma) Substate
This substate occurs only in procedures where both platelets and plasma are collected. The plasma valve
is commanded to the collect position and the plasma pump speed changes to regulate the RBC line
hematocrit. The AC and inlet pump speeds change to keep the AC infusion rate at the run configured
value. After enough inlet volume has been processed to stabilize the LRS chamber, the Trima Accel
system advances to the Platelet Plasma substate or the Pre-Platelet No Plasma substate.
Table 3-31: Pre-Platelet Plasma Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Collect
Return
3000
Exit Condition
When the target (donor-specific) volume is processed by the inlet pump.
Software Description
3-26
Trima Accel
®
Automated Blood Collection System • Service Manual
Summary of Contents for Trima Accel
Page 1: ...Trima Accel Automated Blood Collection System Service Manual...
Page 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Page 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Page 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Page 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Page 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Page 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Page 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Page 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Page 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Page 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Page 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Page 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Page 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...