Draw/Return Pressure Sensor
Table 6-19: Trima Accel system draw/return pressure sensor specifications
Characteristics
Performance
Conditions
Operating range
−300 mmHg to +500 mmHg
Default alarm point
−250 mmHg ± 30 mmHg (draw)
+310 mmHg ± 37 mmHg (return)
Accuracy
12% of reading or ± 20 mmHg, whichever
is greater
Draw/Return Pressure Sensor
Trima Accel
®
Automated Blood Collection System • Service Manual
6-17
Summary of Contents for Trima Accel
Page 1: ...Trima Accel Automated Blood Collection System Service Manual...
Page 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Page 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Page 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Page 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Page 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Page 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Page 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Page 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Page 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Page 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Page 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Page 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Page 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...