Saline Run
Ensure that all panels and doors are in place before starting the saline run.
1.
Device setup
a.
Allow the Trima Accel device to boot normally.
b.
Verify that there are no alarms after the Trima Accel device has fully booted.
2.
Load a disposable set
a.
Load a disposable set or an appropriate test set for the saline run.
b.
Touch the
Donor Info
button on the main screen and enter the following donor information:
Select gender
Male
Enter the height
6 ft 6 in (198 cm)
Enter the weight
250 lb (114 kg)
Blood type
A+
Hematocrit/hemoglobin
45% or 15.0 g/dL
Platelet precount
200 (× 1,000)
c.
Touch the
Confirm
button to confirm the information and continue.
d.
Select and confirm a product procedure for the type of disposable set used with this saline run.
e.
Touch the
Load System
button.
f.
Enter or accept the REF # of the disposable, if applicable. If “The REF does not match the
selected procedure” is displayed, go back to the Donor Info tab and select a different
procedure.
g.
Follow the on-screen instructions to load the disposable set.
h.
Confirm the donor information, if applicable.
i.
Connect either an anticoagulant bag or a saline bag in place of anticoagulant when prompted
to connect the anticoagulant (AC) bag.
j.
Verify that the AC prime test passes without any alarms after the Priming AC screen appears.
k.
Touch either the
Begin Donor Prep
button or the
Prepare Donor
button and follow the
instructions on the screen. The device uses the saline or AC bag as the donor.
l.
If using a new disposable set, connect the donor needle line to the saline or AC bag, and leave
the white pinch clamp on the sample bag line closed (ignore the displayed instruction to take a
sample). If necessary, use tape to secure the needle and keep it from slipping out of the bag.
m.
If using a test set, connect fluid bag as appropriate for the test set.
n.
Touch the
Start Draw
button.
o.
The Blood Prime In Progress screen is displayed. Wait for the device to prime.
p.
The Draw In Progress screen is displayed.
q.
Verify that the device completes the first draw and return cycle without alarms.
Maintenance and Calibration
5-14
Trima Accel
®
Automated Blood Collection System • Service Manual
Summary of Contents for Trima Accel
Page 1: ...Trima Accel Automated Blood Collection System Service Manual...
Page 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Page 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Page 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Page 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Page 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Page 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Page 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Page 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Page 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Page 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Page 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Page 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Page 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...