Positioning the Linear Actuator Sensors
The linear actuator sensors may need to be repositioned if there are cassette alarms or if the machine fails
the cassette plate test in Autotest.
The panel on the same side of the device as the IV pole must be removed.
1.
Boot the device to Service mode and connect with the service software.
2.
Enable the 24 V switch.
3.
Raise the cassette plate.
4.
The two optical sensors are mounted below the pump panel on one of the square members that
house the linear actuator guides. Loosen the lower accessible screw for the top optical sensor.
Figure 5-4: The lower accessible screw (circled)
5.
Place a 0.035 in (0.89 mm) or 0.036 in (0.91 mm) feeler gauge under the upper left corner of the
cassette plate.
6.
Command the cassette plate down and verify that the service software cassette position indicator
status is
Unknown
.
•
If the service software cassette position indicator status is
Down
•
If the service software cassette position indicator status is
Up
•
If the service software cassette position indicator status is
Unknown
Maintenance and Calibration
5-8
Trima Accel
®
Automated Blood Collection System • Service Manual
Summary of Contents for Trima Accel
Page 1: ...Trima Accel Automated Blood Collection System Service Manual...
Page 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Page 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Page 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Page 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Page 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Page 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Page 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Page 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Page 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Page 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Page 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Page 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Page 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...