occupies the eighth bit of data, so that depending on the communications software used, the data length may have
to be changed to seven bits (as the parity bit is then added after the seventh bit).
The default data rate on a new unit is 9600 baud, but may be changed by way of the comm. interface to 300, 600,
2400, 4800, or 9600 baud.
A description of the various commands follows. They are separated into the different categories. First is presented
a Command Summary followed by a detailed description of the commands.
606
COMMAND SUMMARY - FOR SOFTWARE VERSIONS 1.19 - 1.20
NOTE: All commands that do not require a response otherwise, respond with an "A" and a carriage return if
properly recognized. Whenever any command is not properly recognized, the Modular unit will respond with a
question mark, "?" followed by a carriage return. All other command responses end with a carriage return.
Commands which are not stored using the "SE<cr>" command will be lost when powder is interrupted and unit is
repowered.
COMMAND
RESPONSE
"B" MODE
BF<cr>
Exit "B" Mode (can be stored).
BN<cr>
Go to "B" Mode (can be stored).
BO<cr>
Output Data in "B" Mode.
COLD CATHODE
AC<cr>
Auto Mode for Cold Cathode Gauges.
EC<cr>
Turn Cold Cathode Gauges "ON".
NC<cr>
Turn Cold Cathode Gauges "OFF".
COMMUNICATIONS (Can be stored)
BE<cr>
Blank echo.
EE<cr>
Enable echo.
PE<cr>
Insert and test for even parity.
PF<cr>
Turn off parity check and insertion.
PO<cr>
Insert and test for odd parity.
SBll<cr>
Set baud rate. (See Table in Section 603).
SE<cr>
Store entries.
CONFIGURATION
AR<cr>
Outputs relay configuration.
DI<cr>
Modular unit performs internal diagnostics.
Faults will be reported as " P5LO ..."
If no faults "No Err ".
Sx<cr>
Returns module type installed for Station x.
SV<cr>
Send software version.