INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:[email protected] web site:www.televac.com
MM-200_im REV M
Page 48 of 160
TELEVAC
MM200 0-10V OUTPUT FORMATS
1E & 1F DIAPHRAGM GAUGES
RS232
CODE
display
abbr
NAME
R1EN
Linr
LINEAR
0-1000 TORR/10BAR
0-10V
10V =1000 TORR/10BAR----------
STD
*
LIN/DEC
“
2.0V/DEC
“
RIEL
LoG3
LOG
“
3.33V/DEC
“
2A THERMOCOUPLE GAUGES
R2AN
Ln1t
LINEAR
0-1000 MICRONS
0-10V
10V =1000 MICRONS---------------
STD
R2AL
LoG3
LOG
0-1000 MICRONS
3.33V/DEC
10V =1000 MICRONS
R2AT
2Atr
2A TRAD’L
0-20,000 MICRONS
0-10V
10V = 0 MICRONS (RL)
R2AC
2Ctr
2C TRAD’L
0-20,000 MICRONS
0-10V
10V = 0 MICRONS (RL)
R2AR
-Lin
LIN REV’D
1000 MICRONS-0
0-10V
10V = 0 MICRONS (RL)
3F, 3D & 3E HOT CATHODE GAUGES
R3DM
L-d1
LIN/DEC
10-11 TO 10-2 TORR
1.00V/DEC
10V = 10-2 TORR---------------------
STD
R3DL
LG10
LOG
“
1.00V/DEC
10V = 10-2 TORR
R3Dx
Ln-x
LINEAR
0-10-x TORR
0-10V10v =
10-X
(8=,.X.,=2)
10V = 10-x (8=,>x>,=2)
R3FM
L-d1
LIN/DEC
10-11 TO 10-2 TORR
1.00V/DEC
9V =10-2 TORR -
---------------------
STD
R3FL
LG10
LOG
1.00V/DEC
9V =10-2 TORR
R3Fx
Ln-x
LINEAR
0-10-x TORR
0-10V
10V = 10-x(8=,>.x>,=2)
4A CONVECTION GAUGES
R4AN
L-d1
LIN/DEC
10-3 TO 1000 TORR
1.00V/DEC
10V = 1000 TORR--------------------
STD
R4AU
Ln1t
LINEAR
0-1000 MICRONS
0-10V
10V =1000 MICRONS
R4AT
Ln10
LINEAR
0-10 TORR
0-10V
10V = 10 TORR
R4AL
LoG6
LOG
10-3 TO 1000 TORR
1.67V/DEC
10V = 1000 TORR
JMPR
4A NON-LIN
10-3 TO 1000 TORR
0-10V
10V = 760 TORR
*
2A TRAD’L
0-1000 MICRONS
0-10V
10V = 0 MICRONS (RL)
*
2C TRAD’L
0-20,000 MICRONS
0-10V
10V = 0 MICRONS (RL)
5A,B,C,D,E,F CAP DIAPHRAGM GAUGES
R5AN
Linr
LINEAR
0-FULL SCALE
0-10V
10V = FS--------------------------------
STD
R5AL
LoG3
LOG
3 DECADES
3.33V/DEC
10V = FS
5A FS = 1000 TORR 5D FS = 1
TORR
5B FS = 100 TORR 5E FS = 0.1
TORR
5C FS = 10 TORR 5F FS =
SPECIAL
7B COLD CATHODE GAUGE MODULE (2-6200-227)
R7BM
L-d2
LIN/DEC
10-7 TO 10-3 TORR
2.00V/DEC
10V = 10-3 TORR---------------------
STD
R7BL
LoG3
LOG
10-6 TO 10-3 TORR
3.33V/DEC
10V = 10-3 TORR
R7BE
LoG4
EXT’D LOG
10-7 TO 10-3 TORR
2.50V/DEC
10V = 10-3 TORR
R7BN
Linr
LINEAR
10-6 TO 10-3 TORR
0-10V
10V = 10-3 TORR
R7BA
7Atr
7ACTRADL
10-5 TO 10-3 TORR
0-10V
10V = 10-3 TORR
R7BT
B7b
B7B2 TRADL
10-7 TO 10-3 TORR
0-10V
10V = 10-3 TORR
7E COLD CATHODE GAUGE
R7EM
L-d1
LIN/DEC
10-8 TO 10-2 TORR
1.00V/DEC
10V = 10-2 TORR (4-10V)----------
STD
R7EL
LoG6
LOG
10-8 TO 10-2 TORR
1.67V/DEC
10V = 10-2 TORR (0-10V)
R7Fx
Ln-x
LINEAR
0 TO 10-x TORR
0-10V
10V = 10-x TORR (7=, >x>,=2)
7F (7FC, 7FCS) COLD CATHODE GAUGE
R7FM
L-d1
LIN/DEC
10-11 TO 10-2 TORR
1.00V/DEC
9V = 10-2 TORR-----------------------
STD
R7FL
LG10
LOG
10-11 TO 10-2 TORR
1.00V/DEC
9V = 10-2 TORR
R7x
Ln-x
LINEAR
0 TO 10-x TORR
0-10V
10V = 10-x TORR (8=,>x>, =2)
Note: “M” is displayed as “ii”
RECUT205 110598
SEND “SE” AFTER SENDING NEW FORMAT COMMAND
TO RETAIN IT AFTER POWER OUTAGE
Table 4.1