702 VACUUM CALIBRATION OF VACUUM MEASURING INSTRUMENTS ON VACUUM
CHAMBERS
A Televac instrument with tubes and cables will be given an NIST traceable calibration and so certified by Televac.
This instrument will be used as a "master gauge" for re-calibration of the installed vacuum gauges on a periodic
basis.
[When practical, the master gauge can be used as a spare in an emergency to assure continuity of production.]
EQUIPMENT REQUIRED
In order to use the On-site Calibration Procedure, the following equipment is required.
A. Vacuum Chamber Equipment
Each vacuum chamber should be fitted with an instrument with special features. The MM200 meets
these requirements.
1. A "zero adjust" ("heater adjust") calibration potentiometer for the thermocouple gauge(s), located
on the rear panel.
2. A 10
-3
Torr calibration potentiometer for the cold cathode gauge(s), located on the rear panel.
3. Cold cathode gauge tubes of the latest improved design. (Retrofit existing tubes to the dual
magnet configuration or replace, if necessary.
B. Master Gauge
A master gauge is an instrument with the same features as above. This gauge will have one cold
cathode gauge tube with a 50' cable and at least one thermocouple gauge tube with a 50' cable.
The instrument, tubes and cables will have NIST traceable calibration performed by Televac for the
cold cathode station and for thermocouple station(s)
This master gauge with its tubes and cables should be returned to Televac annually for an NIST
traceable calibration. When external calibration potentiometers are provided they must be sealed
after the calibration is completed.