Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
33
4.
Removing Rear Case
i.
Remove the handle from the rear of the GC by unscrewing
the 4 knobs and lifting handle free.
ii.
Unscrew 6 screws on the back of the GC and remove the
rails the screws were in.
iii.
Carefully lift the rear case off of the GC by gently pulling
upwards while the GC rests on its front. Take care not to
pull any tubes loose from around the inlet manifold.
Note: If the case fan attached to the back of the GC enclosure
has knurled thumb screws, remove all four thumb screws
and unplug and remove the fan before taking the case off.