5.4 FID Removal
1.
Disconnecting the Flow Restrictor
i.
Using a 1/4” and a 3/16” wrench on the fitting for the flow
restrictor, disconnect the nut from the FID inlet fitting and
gently remove the flow restrictor.
2.
Remove hydrogen and air lines from the FID.
i.
Using a ¼” wrench unscrew both the hydrogen and air
lines connecting the FID manifolds to the GC.
ii.
The coaxial cable to the collector may be disconnected to
aid removal.