Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
47
20.
Replace the front on the unit using the 4 screws removed in Step
1.
Installation is complete. After powering up the instrument, all of the
column's information can be viewed either from the touchscreen display
in the “Column Module” screen or in under the “GC Info” tab of the
“Special Features” box in Chromperfect.
5.6 Replacing the Glow Plug
Caution:
If the unit is on and at operating temperature the
glow plug will be very hot. It should NOT be
touched without proper precautions. If the unit is
off and near room temperature the glow plug will
be at a safe temperature and can be touched
without tools or gloves.
1.
Remove the red wire that connects the glow plug to the FID
board near the base of the FID. This can be done by lightly
pulling on the rubber sleeve covering the metal end of the glow
plug wire.
Summary of Contents for FALCON GC
Page 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Page 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Page 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Page 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Page 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Page 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Page 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Page 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Page 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Page 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...