Maintenance & Troubleshooting
Falcon GC
24
Teledyne Analytical Instruments
5.1 Troubleshooting
Columns
Problem:
Each column, if operated within the proper temperature
limits, should have a lifetime of 1 year+. TAI
recommends having spare columns on hand in case of
emergencies. Over time you may experience a changes
in the column phase or column hardware.
Effects:
Tailing peaks
Fronting peaks
Retention time shift
Runaway temperature or faulty temperature reading
(hardware failure)
Solution:
The column may need to be replaced. Please refer to
the column change guide if that is the case. In the case
of tailing, fronting, or retention time shifting it may be
possible to flush the column with solvent to clear out
any pollutants or “heavy” material. This should be
done while the column has been removed from the GC.
In the case of a hardware failure, or persistent poor
performance, you will have to ship the column back to
Teledyne for refurbishment
Detectors (FID)
Problem:
The FID may fail to light or the flame may go out due
to fluctuations in air or H2 pressures
Effects:
Chromatogram is produced with no peaks on the FID
channel
Solution:
Turn off the Hydrogen and Air to the detector and
repeat the “Ignite” process. You should hear a single
pop. Multiple pops may indicate that the flame is
lighting and extinguishing alternately. You can check
Summary of Contents for FALCON GC
Page 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Page 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Page 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Page 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Page 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Page 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Page 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Page 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Page 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Page 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...