TROUBLESHOOTING and REPAIR
87
Table 11-4. Temperature Controller – Secondary Parameter Settings
DISPLAY
PARAMETER
NAME
SET TO
COMMENTS
OFF
0
Sets the controller to reverse action (hotter
temperatures = less heating; lower temperatures =
more heating) and turns off the heater should the
thermocouple input fail.
Input type
(Refer to
COMMENTS,
this row)
You may have one of three thermocouple Types; the
wire colors distinguish which one you have:
red & black (Type S): set to 6
red & yellow (Type K): set to 3
red & orange (Type N): set to 12
If replacing the thermocouple, carefully follow the
instructions provided in Section 11.5.6.
Digital Filter
5
Sets the response time of the digital noise filter to 5
sec.
Lower input range
limit
40
Sets the lower end of the controller full scale bandwidth
to 40˚C
Upper input range
limit
1100
Sets the upper end of the controller full scale
bandwidth to 1100˚C
High Alarm Type
10
Sets the High alarm point to trigger when the process
value exceeds the high alarm set point only after the
process value rises above the lower limit for the first
time after power up
Low Alarm Type
10
Sets the Low alarm point to trigger when the process
value exceeds the high alarm set point only after the
process value rises above the lower limit for the first
time after power up
Alarm Hysteresis
3
Sets the hysteresis bandwidth for both alarm points to
3% of full scale.
Decimal Point
0
Sets the decimal point for PV & SV readings to NONE
Process variable
Offset
0
OFF
Set variable
Offset
0
OFF
Celsius/Fahrenheit
C
Sets controller to perform function on Celsius
temperature scale.
Fuzzy Logic
ON
Turns of the controllers advanced fuzz logic feature
which further reduces overshoot and fluctuations.
Ignore any other parameters that may appear.
11.7.
Technical Assistance
If this addendum and its trouble-shooting / repair sections do not solve your problems,
technical assistance may be obtained from TAPI, Technical Support, 9970 Carroll
Canyon Road, San Diego, CA 92131. Phone: +1 858 657 9800 or 1-800 324 5190. Fax:
+1 858 657 9816. Email:
.
Before you contact Technical Support, fill out the problem report form in Appendix C,
which is also available online for electronic submission at http://www.teledyne-
api.com/forms/index.asp.
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......