TROUBLESHOOTING and REPAIR
Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
68
CAUTION
Make sure the M501-TRS is turned off and no AC power is being
supplied to the heater before proceeding.
•
Check the resistance across the heater coil. It should be
approximately 16.5 ohms.
•
Check the wiring between the heater and the power supply.
•
Check to make sure the M501-TRS cooling fan is operating properly
and that the instrument is properly ventilated. Poor ventilation can
cause the M501-TRS’ thermal switch to turn off the converter to
prevent overheating.
•
Check to make sure that the thermal switch and heater control relay
are operating properly.
11.3.
Other Performance Problems
11.3.1.
Excessive noise
In addition to the causes listed in the
Excessive noise
section in the T101 Manual (PN
07266), an excessively noisy TRS measurement can be caused by hysteresis or
fluctuations in the temperature of the TRS converter oven in the M501-TRS. If this is
the case, perform the autotune procedure described in Section 7.4. of this addendum.
11.4.
Subsystem Checkout
In addition to the information contained in the
Subsystem Checkout
section in the T101
Manual - PN 07266, the following diagnostic procedures are useful for troubleshooting
and diagnosing problems with your T102 and M501-TRS.
11.4.1.
Checking the Efficiency of the M501-TRS SO
2
Scrubber
See Section 9.1.1.2 of this addendum
11.4.2.
Checking the Efficiency of the M501-TRS TRS
SO
2
Converter
To check to see if your TRS
SO
2
converter is operating properly:
1.
Set the analyzer to TRS measurement mode (see the Setting the Gas
Measurement section in the T101 Manual - PN 07266).
2.
Bypass the scrubber inside the M501-TRS.
a.
Unscrew the pneumatic fittings from the scrubber.
b.
Connect them with a stainless steel or Teflon
®
union.
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......