SPECIFICATIONS AND APPROVALS
Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
16
•
Range: Any range from 50 parts per billion (ppb) to 10 parts per million
(ppm).
•
Ambient temperature range of 5
o
C to 40
o
C.
•
Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
•
Sample filter: Equipped with PTFE filter element in the internal filter
assembly.
•
Sample flow of 650 +/- 65 cc/min.
•
Vacuum pump (internal or external) capable of 14"Hg Absolute pressure
@ 1 slpm or better.
•
Software settings:
Dynamic span
OFF
Dynamic zero
OFF
Dilution factor
OFF
AutoCal
ON or OFF
Dual range
ON or OFF
Auto-range
ON or OFF
Temp/Pressure compensation ON
Under the designation, the analyzer may be operated with or without the following
optional equipment:
•
Rack mount with chassis slides.
•
Rack mount without slides, ears only.
•
Zero/span valve options.
•
Internal zero/span (IZS) option with either:
•
SO2 permeation tube - 0.4 ppm at 0.7 liter per minute;
certified/uncertified.
•
SO2 permeation tube - 0.8 ppm at 0.7 liter per minute;
certified/uncertified. Under the designation, the IZS option cannot be
used as the source of calibration.
•
4-20mA isolated analog outputs.
•
Status outputs.
•
Control inputs.
•
RS-232 output.
•
Ethernet output.
•
Zero air scrubber.
•
4-20mA, isolated output.
2.3.
CE Mark Compliance
See Section titled
CE Mark Compliance
in the T101 Manual - PN 07266
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......