15
2.
SPECIFICATIONS AND APPROVALS
2.1.
Specifications
As there are no significant differences between the performance specifications for the
T102 and the T101 please refer to in the Specification section of the T101 Manual - PN
07266. However, the AC power specifications for the T102 are as follows:
T102 AC Power
: 100V – 120V, 60Hz (200W); 220V – 240V, 50Hz (211W)
2.1.1.
M501-TRS Specifications
Table 2-1: Model 501 Basic Unit Specifications
Minimum Converter Efficiency
H
2
S >95%
COS >90%
CS
2
>90%
Maximum TS Concentration for
specified conversion efficiency
20 ppmv
Sample Flow Rate
650cc/min. ±10% - driven by T102 pneumatic system
Optimum Converter
Temperature
850
°
C (factory setup)
Maximum Converter
Temperature
1100
°
C
Dimensions H x W x D
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
16 lbs (7.3 kg)
26 lbs (11.8 kg) CE version
AC Power Rating
115 V, 50/60 Hz - 400 Watts;
230 V, 50/60 Hz - 575 Watts; CE Version
Internal Alarms
High Alarm Point: 900
°
C
Low Alarm Point:
800
°
C
Alarm Output Relay
SPST - 1 point: Alarm output is energized should either the temperature
controller’s high or low internal alarm set points be activated.
Alarm Output Rating
220V AC/30V DC, 1A (resistive load)
Environmental
Installation category (over-voltage category) II; Pollution degree 2
Certifications
IEC 1010-1 / 61010-1:93 (includes A1) + A2:95,
For indoor use at altitudes ≤ 2000m only
2.2.
EPA Equivalency Designation
No EPA equivalency standards exist for TRS measurement, however, the T102 analyzer
qualifies for EPA equivalency designation as Reference Method Number EQSA-0495-
100 per 40 CFR Part 53 when operated under the following conditions:
•
Measurement Mode: SO
2
single gas mode.
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......