51
9.
INSTRUMENT MAINTENANCE
The following table supersedes the T101 Preventative Maintenance Table in the Maintenance Schedule section in the T101 Manual - PN 07266.
Table 8-1. T102 Preventive Maintenance Schedule
ITEM
ACTION
FREQUENCY
CAL
CHECK
T101 MANUAL
SECTION
DATE PERFORMED
M501 SO2 scrubber
Replace scrubber
material
As required
Yes
9.1.1.3 of this
addendum
1
Particulate filter
Change particle filter
Weekly
No
9.3.1
Verify test functions
Review and evaluate
Weekly
No
Appendix C
Zero/span check
Evaluate offset and
slope
Weekly
--
7.3, 7.6, 7.9
1
Zero/span
calibration
Zero and span
calibration
Every 3 months
--
7.2, 7.4, 7.5, 7.7,
7,8
1
External zero air
scrubber (option)
Exchange chemical
Every 3 months
No
9.3.4
1
Perform flow check
Check Flow
Every 6 Months
No
11.5.2
1
Sample chamber
optics
Clean windows and
filters
Annually or as
necessary
Yes
9.3.6
1
Critical flow orifice
& sintered filters
Replace
Annually
Yes
9.3.7
Internal IZS
Permeation Tube
Replace
Annually
YES
9.3.2
Perform pneumatic
leak check
Verify Leak Tight
Annually or after
repairs involving
pneumatics
Yes
11.5.1
2
Pump diaphragm
Replace
Every 2 years, or as
necessary
Yes
See instruction in
diaphragm kit
PMT sensor
hardware calibration
Low-level hardware
calibration
On PMT/ preamp
changes if 0.7 < SLOPE
or SLOPE >1.3
Yes
11.5.4
Desiccant bags
Replace
Any time sensor assy
inspection plate is opened
(reference Section 11.5.3)
1
These Items are required to maintain full warranty, all other items are strongly recommended.
2
A pump rebuild kit is available from TAPI Technical Support including all instructions and required parts (see Appendix B for part numbers).
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......